Main Parameters:
Parameters | Specifications | |
Spectral range | 600 ~1700nm | |
Scan span | 0.2 ~1100nm(Full range span) | |
Wavelength accuracy | ±0.02nm( 1520 ~1620nm)、 ±0.04nm( 1450 ~1 520nm)、 ±0.10nm(Full range span) | |
Wavelength linearity | ±0.01nm( 1520 ~1580nm) | |
Wavelength repeatability | ±0.005nm( 1min) | |
Wavelength resolution setting | 0.02、0.05、0.1、0.2、0.5、1、2nm | |
Min. Sampling resolution | 0.001nm | |
Number of sampling points | 101~50001、AUTO | |
Power sensitivity setting | NORMAL、MID、HIGH1、HIGH2 and HIGH3 | |
Power sensitivity | -90dBm( 1300-1620nm)、-85dBm( 1000 ~1300 nm)、-60dBm( 600-1000nm)(Sensitivity :HIGH3) | |
Max. input power | +20dBm(Each channel, full wavelength range) | |
Max. safe input power | +25dBm(Total input power) | |
Power accuracy | ±0.4dB(1310/1550nm,Input Power:-20dBm,Sensitivity: MID) | |
Power linearity | ±0.05dB(Input Power :-50~ +10dBm) | |
Power flatness | ±0.1dB(1520 ~1580nm)、±0.2dB(1450 ~1520nm、 1580 ~1620nm) | |
Polarization dependence | ±0.08dB(1550nm) | |
Dynamic Range | Resolution :0.02nm | 63dB(Peak±0.2nm ,Typ.66dB)、46dB(Peak±0.1n m ,Typ.50dB) |
Resolution:0.05nm | 73dB(Peak±1.0nm ,Typ.76dB)、70dB(Peak±0.4n m ,Typ.73dB)、63dB(Peak±0.2nm ,Typ.66dB) | |
Resolution:0.1nm | 67dB(Peak±0.4nm ,Typ.70dB)、57dB(Peak±0.2n m ,Typ.60dB) | |
Stray light suppression rate | 76dB | |
Optical return loss | 35dB(When using the APC connector) | |
Applicable optical fiber | SM(9.5/125μm)、GI(50/125μm、62.5/125μm)、 Large core fiber( Max. 200μm) | |
Light source output | Factory installed standard:DFB light source(1550±5nm); Replaceable light source: see options for details | |
Display | 12.1-inch touchscreen | |
Storage | 128GB | |
Interfaces | USB/Ethernet/RS232C/VGA | |
Working conditions | Ambient temperature :0℃ ~ 40℃ ; Humidity: ≤80% Performance guaranteed temperature:18~28℃ | |
Dimensions | W×H ×D =426mm×221mm×450mm | |
Weight | About 19kg | |
Power supply | 100~240VAC、50\60Hz | |
Max. power consumption | About 100W | |
Warm-up time | At least 1h (After warming up, the built-in light source is required for built-in light source calibration.) |
Note: Wavelength calibration and optical axis alignment can be performed using an external light source.
Appendix 1: Model and Part Number Comparison Table
Part Number | Stock Number | Title | Spectral Range | Span | Wavelength Accuracy | Wavelength Repeatability | Wavelength Resolution Setting | Min.Sampling resolution |
6362D-V 1.0 | 6362D Spectrometer | 600~170 0nm | 0.2nm~11 00nm(Full range span) | ±0.02nm(1 520~1620 nm)、 ±0.0 4nm(1450 ~1520nm)、 ±0.10 nm(Full range) | ±0.005nm(1 Min.) | 0.02、0.05、 0.1、0.2、0.5、 1 and 2nm | 0.001nm |